Use Instructions
Vital Signs Monitor User Manual Monitoring NIBP
- 55 -
14.9 Setting Inflation Value
To change the inflation value:
1. Select Menu > Module Switch > Inflation value;
2. Choose AUTO or other inflation values from the pull-down list.
If other inflation values are chosen, the preset value by users will be adopted as the
inflation value when measuring blood pressure.
If AUTO is chosen, the default value will be adopted as the inflation value when
measuring blood pressure.
14.10 Manometer Mode
The device supports manometer function to check whether the relative pressure coefficient in the
NIBP measurement is accurate.
Access Menu > User Maintain > Manometer. The device will display the pressure value in real
time and prompt ‗pressure testing‘.
To exit, click the Back symbol < on top left area.
14.11 Cleaning Mode
The device supports cleaning mode for NIBP dust removal maintenance, which is applicable to
adult and under the precondition that the cuff is removed from the device.
Access Menu > User Maintain > Cleaning Mode. The device will display the pressure value in
real time, prompt ‗cleaning‘, and display 3-minute countdown. During the cleaning, user can use
dry cotton swab to clean NIBP port.
To exit, click the Back symbol < on top left area.