Use Instructions

Vital Signs Monitor User Manual Monitoring NIBP
- 53 -
14.5.1 Operation Prompts
Slide NIBP parameter area left to select and set:
1. Manual Measuring
Click or press the button on the front panel to start measurement.
2. Continuous measurement
Click to start measurement. The continuous measurement will last 5 minutes with a fix
interval of 5 seconds.
3. AVG measurement
Click , set the interval and times, and click Confirm to start measurement.
4. Stopping measurement
During measurement, click or press the button on the front panel at any time to
stop measurement.
For continuous and AVG measurement, NIBP parameter area will display the countdown time,
and the measurement will continue after the countdown. To exit the whole measurement mode,
click the corresponding measurement icon , again.
14.5.2 Correcting the Measurement if Limb is not at Heart Level
To correct the measurement if the limb is not at heart level to the displayed value:
Add 0.75 mmHg (0.10 kPa) for each centimeter
higher or
Deduct 0.75 mmHg (0.10 kPa) for each
centimeter lower or
Add 1.9 mmHg (0.25 kPa) for each inch higher
Deduct 1.9 mmHg (0.25 kPa) for each inch
lower
14.6 Resetting NIBP
When the pressure does not work properly and the system fails to give a message for the problem.
Please click Menu > User Maintain > Reset to activate self-test procedure, and thus restore the
system from abnormal performance.