Use Instructions
Vital Signs Monitor User Manual Prompt Information
- 39 -
Message
Cause
Action Taken
Battery1 Charge
Voltage Too High
The charging voltage in
battery of the monitor is too
high
Please replace the battery.
Device Overheated
The temperature of the
monitor is too high.
Turn off the monitor in time and
restart it after cooling.
Recorder Out Of Paper
Recorder Out Of Paper
Please install the paper.
Recorder Setup
Needed
The user presses the Record
button when Recorder is not
configured.
Notify the manufacturer‘s service
staff to install and set the
recorder.
10.3 Operation Prompt
Message
Cause
SpO
2
Searching Pulse
SpO
2
module is analyzing the patient signal and searching
for the pulse to compute the saturation, when sensor is
connected with patient.
Please Start
NIBP module is in idle status.
Measuring
During NIBP measurement.
Measurem. Canceled
During NIBP measurement, user presses the hardkey of
Start/stop NIBP measurement.
Done
NIBP measurement is completed.
Source: SpO
2
or NIBP
PR source is displayed in parameters area.
No Source
Both SpO
2
and NIBP are off. PR has no source.
Check whether the airway is
connected to the atmosphere
1. In NIBP cleaning mode, if pressure keeps over 15
mmHg for more than or equal to 30 s±2 s;
2. In NIBP cleaning mode, if pressure is more than or
equal to 140 mmHg±3 mmHg;
Measure time out
(F3000 TEMP)
No measuring result after the module entering Predict
state for 30 s.
Probe TEMP Too High
(F3000 TEMP)
The initial temperature of the probe is at >+33 ℃ and ≤
+40 ℃.
TEMP Is Warming Up
(F3000 TEMP)
The monitor displays this message after taking the sensor
out of the bracket and warm-up is in process.
Predict Over
(F3000 TEMP)
After the Predict measuring is over, the data and message
display on the interface.