Use Instructions

Vital Signs Monitor User Manual Prompt Information
- 38 -
Message
Cause
Action Taken
TEMP Error P6
Unable to pre-heat probe tip.
Stop using measuring function of
TEMP module; notify biomedical
engineer or Manufacturer‘s
service staff.
NOTE: Measure readings
displayed on the screen are
unreliable when the monitor
indicates Temp Error P06.
Temp COMM Fail
TEMP module failure or
communication failure.
Stop using measuring function of
TEMP module; notify biomedical
engineer or Manufacturer‘s
service staff.
No TEMP Sensor
Probe configuration (or no
probe connected) error.
Well connect the probe and the
monitor, and measure again.
Measure Site Error
The probe in use is not
consistent with the measure
position set on the monitor.
Correctly set the measure position
on the monitor.
F3000 TEMP /Internal forehead TEMP
TEMP Exceed Limit
F3000 TEMP: The TEMP
value is out of the range of
+30 C ~ +43 C.
Internal forehead TEMP: The
TEMP value is out of the
range of +15.5 C ~ +43.3 C.
F3000 TEMP: Put the probe into
the probe well; take it out and
measure again.
Internal forehead TEMP: Check
the integrity of the probe, make
sure it is clean, and take a new
measurement.
Others
Battery Low
The battery power of the
monitor or the Stand is Low.
Please change the battery or
charging.
Battery1 Error
Malfunction in Battery of the
monitor
Replace the battery and restart the
monitor. If the problem persists,
notify the manufacturer‘s service
staff.
Battery 1 Overheated
The temperature in battery of
the monitor is too high.
Turn off the monitor in time and
restart it after cooling.