Use Instructions
Vital Signs Monitor User Manual Prompt Information
- 37 -
Message
Cause
Action Taken
NIBP Interference
Signal noise is too large or
pulse rate is not regular due to
the patient movement.
Make sure that the patient under
monitoring is motionless.
NIBP Leak Test Error
During leak testing, deflation is
failed.
Check the connections and the
wrapped cuff to see whether they
are all prepared well.
If failure persists, please notify
biomedical engineer or
manufacturer‘s service staff.
F3000 TEMP
TEMP Error E1
System error during
synchronization.
Stop using measuring function of
TEMP module; notify biomedical
engineer or Manufacturer‘s
service staff.
TEMP Error E2
System error during patient
thermistor calibration.
TEMP Error E3
System error during heater
thermistor calibration.
TEMP Error E4
System timing error.
Probe Heater Error
Heater error.
TEMP Error P2
Monitor Mode patient
thermistor unstable or out of
range.
TEMP Error P3
Monitor Mode heater
thermistor unstable or out of
range.
TEMP Error P4
Predict Mode patient
thermistor unstable or out of
range.
TEMP Error P5
Predict Mode heater thermistor
unstable or out of range.