Use Instructions
Vital Signs Monitor User Manual Prompt Information
- 36 -
Message
Cause
Action Taken
NIBP
NIBP Leak
NIBP pump, valve, cuff or tube
has a leakage.
Check the connections and the
wrapped cuff to see whether they
are all prepared well.
If failure persists, please notify
biomedical engineer or
manufacturer‘s service staff.
NIBP Over Pressure
Pressure has exceeded the
specified upper safety limit.
Measure again, if failure persists,
stop measuring function of NIBP
module and notify biomedical
engineer or manufacturer‘s
service staff.
NIBP Init Pressure
High
The initial pressure is too high
during measuring
NIBP Aux Excessive
Pressure
Pressure has exceeded the
second safety limit as
specified.
Notify biomedical engineer or
manufacturer‘s service staff.
NIBP Time Out
Measuring time has exceeded
the specified time.
Measure again or use other
measuring method.
NIBP Self Test Error
Sensor or other hardware
errors.
Measure again, if failure persists,
stop using measuring function of
NIBP module and notify
biomedical engineer or
manufacturer‘s service staff.
NIBP Cuff Type Error
The cuff type used isn‘t
consistent with the patient
type.
Confirm the patient type and
change the cuff.
NIBP
SystemPress.Abnormal
Atmospheric pressure or
system pressure is abnormal.
The valve is occluded so that
deflation is failed.
Check whether the airway is
occluded or pressure sensor
works properly. If the problem
still exists, contact your service
personnel.
NIBP System Failure
NIBP is not calibrated.
Contact your service personnel.
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Use other methods to measure
blood pressure.
NIBP Range Exceeded
Maybe the patient blood
pressure value is beyond the
measurement range.
Use other methods to measure
blood pressure.
NIBP Loose Cuff
Cuff is not properly wrapped
or no cuff is connected.
Properly wrap the cuff.