Use Instructions
Vital Signs Monitor User Manual Installation
- 21 -
NOTE:
1 Check all the functions of the monitor and make sure that the monitor is in good
status.
2 If rechargeable batteries are provided, charge them after using the device every time,
to ensure the electric power is enough.
3 After long-time continuous running, please restart the monitor to ensure the monitor‟s
steady performance and long lifespan.
3.5 Connecting Sensor to Patient
Connect all the necessary patient sensors between the monitor and the patient.
NOTE:
For information on correct connection, refer to related chapters.
3.6 Checking the Recorder
If your monitor is equipped with a recorder, open the recorder‘s door to check if paper is properly
installed in the slot. If no paper exists, refer to Chapter Recording for details.
3.7 Setting Date and Time
To set the date and time:
1. Select Menu > Date/Time.
2. From left to right: year, month, day, hour, minute, second. Click the icon on the upper
right area to confirm the operation.
NOTE:
1 If the system is not used for a longer period of time, its system time may be
inaccurate. In this case, readjust the system time after powering on.
2 If the system time cannot be saved and resumes the default value after restart,
contact the service department of EDAN to replace the button cell in main board.
3.8 Handing Over the Monitor
If you are handing over the monitor to the end-users directly after configuration, make sure that it
is in normal working status and let user know the status.
The users must be adequately trained to use the monitor before monitoring a patient. To achieve
this, they should have access to, and read, the following documentation delivered with the
monitor:
User Manual (this book) - for full operating instructions.
Quick Reference Card - for quick reminders during use.