Use Instructions

Vital Signs Monitor User Manual Installation
- 20 -
Chapter 3 Installation
NOTE:
The monitor installations and settings must be configured by the authorized hospital
personnel.
3.1 Initial Inspection
Before unpacking, check the packaging and ensure that there are no signs of mishandling or
damage. If the shipping cartons are damaged, contact the carrier for compensation and package
them again.
Open the package carefully and remove the monitor and accessories. Check that the contents are
complete and that the correct options and accessories have been delivered.
If you have any question, please contact your local supplier.
3.2 Connecting the Power Cable
For Stand part, connection procedure of the AC power line is listed below:
1 Make sure the AC power supply complies with the following specifications: 100 V-240 V~,
50 Hz/60 Hz.
2 Connect the power cord provided with the monitor. Connect the power cord to connector of
the Stand part. Connect the other end of the power cord to a grounded power outlet.
NOTE:
1 Connect the power cable to the socket specialized for hospital use.
2 Only use the power cable supplied by EDAN.
3.3 Mounting the Monitor
If all situations are normal, please place the monitor on a flat, level surface or on a trolley. About
how to install the trolley for the monitor, please refer to Trolley Installation Guide.
WARNING
The safe loads of the trolley are 11 kg. Exceeding the safe load may cause the device to fall.
3.4 Checking the Monitor
Make sure there is no damage on the measurement accessories and cables. Then turn on the
monitor, check whether the monitor can start normally. Make sure the indicator light flashes once
in turn of three colors and a Di tone sound is heard. If failed, please restart the monitor and
check again. If still failed, please contact the service personnel.
WARNING
If any sign of damage is detected, or the monitor displays some error messages, do not
use it on any patient. Contact customer service center immediately.