Use Instructions
Vital Signs Monitor User Manual Product Specification
- 102 -
A.5 PR
Measuring range
Accuracy
Resolution
PR (SpO
2
)
EDAN
25 bpm to 300 bpm
2 bpm
1 bpm
PR (NIBP)
EDAN
40 bpm to 240 bpm
±3 bpm or 3.5%, whichever is
greater
1 bpm
A.6 TEMP
Internal Forehead TEMP
Measuring method
Infrared measurement
Measuring Mode
Indirect measurement
Measuring site
Forehead
☆Measuring range
34 C ~ 43 C
Unit
C, ℉
Resolution
0.1 C
☆Accuracy
34 C ~ 35 C: 0.3 C
35 C ~ 43 C: 0.2 C
Response time
< 2 s
F3000 Module
☆Measuring range
30 C~43 C
Prediction measurement range
35 C~43 C
Low Temp. Mode prediction
measurement range
33 C~43 C
Working temperature
10 C ~ 40 C
Transport and Storage
-20 C ~ 55 C
Sensor type
Oral /axillary /rectal
Resolution
0.1 C
☆Accuracy
Monitoring Mode and Preditive Mode: 0.1 C
Quick Predictive Mode: 0.3 C
Typical measurement time (after
Oral (Quick Predictive Mode): (3~5) s (non-fever temps);