Use Instructions
Vital Signs Monitor User Manual Product Specification
- 101 -
Pre-inflation Pressure
Adult Mode
Default: 160 mmHg
Range: Auto/80/100/120/140/150/160/180/200/220/240
mmHg
Pediatric Mode
Default: 140 mmHg
Range: Auto/80/100/120/140/150/160/180/200 mmHg
Neonatal Mode
Default: 100 mmHg
Range: Auto/60/70/80/100/120 mmHg
A.4 SpO
2
Measuring Range
0% to 100%
Resolution
1%
☆Data Update Period
1 s
☆Accuracy
☆Adult /Pediatric
2% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
☆Neonate
3% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
SpO
2
Storage Interval
In round or spot-checking mode
30 S (default), 1 min, 2 mins, 5 mins
Sensor
Red Light
(6603) nm
Infrared Light
(90510) nm
Emitted Light Energy
< 15 mW
PI
Measuring Range
0-10, invalid PI value is 0.
Resolution
1
NOTE:
The information about wavelength range can be especially useful to clinicians (for
instance, when photodynamic therapy is performed).