i15 Blood Gas and Chemistry Analysis System User Manual Quality Control (QC) Tests
- 82 -
Figure 6-14 Details of Control Test Results
3. View the details.
4. Press Return to go back to the Control Database screen.
6.1.3.3 Exporting/Uploading/Printing Control Test Data
1. Open the control database.
2. Select the desired control test data.
To Do this
Export
Insert a removable disk into the analyzer, and then press Export.
Upload
Press Upload.
Print
Press Print.
NOTE:
If no control test data is selected before pressing Export/Upload/Print, all of the data
stored in the control database will be exported/uploaded/printed.