i15 Blood Gas and Chemistry Analysis System User Manual Patient Analyzing
- 64 -
p
CO
2
(T) 1 ***
p
O
2
(T) 1 >
HCO
3
-
act 1 <
HCO
3
-
std 1
BE(ecf) 1
Ca
++
(7.4) 1
…
Reference Ranges
pH [7.35 - 7.45]
p
CO
2
[35.0 - 45.0] mmHg
p
O
2
[80.0 - 100.0] mmHg
Na
+
[138.0 - 146.0] mmol/L
K
+
[3.50 - 4.90] mmol/L
Ca
++
[1.12 - 1.32] mmol/L
Cl
-
[98 - 109] mmol/L
Glu [3.9 - 5.8] mmol/L
Lac [0.36 - 1.7] mmol/L
Hct [35 – 51] %
Cre [0.6 - 1.3] mg/dL
BUN
5.2.2 Understanding Result Symbols
The following table shows the symbols that may appear on the screen:
Symbol Description
> or < The result is above or below the measurement range.
↑ or ↓ The result is above or below the reference range.
--- The measured parameter fails calibration.
xxx
The measured parameter fails quality control (QC) tests, and QC lockout function is
enabled in Setup.
***
The measured parameter fails quality control (QC) tests, and QC lockout function is
disabled in Setup.
The result for the calculated parameter is valid, but the measured parameter used to
determine this calculated parameter fails quality control (QC) tests, and QC lockout
function is disabled in Setup.
??? The result for the measured parameter is invalid.
NOTE:
Invalid test result for a calculated parameter will not be displayed on the screen.
Valid test result for a calculated parameter will not be displayed on the screen, if the
measured parameter used to determine the result fails quality control (QC) tests, and
QC lockout function is enabled in Setup.