i15 Blood Gas and Chemistry Analysis System User Manual Patient Analyzing
- 56 -
5.2 Patient Analyzing
NOTE:
Ensure that the displayed date and time are correct before each sample test, because
they are part of the patient sample data. Contact the administrator if the date and/or time
are incorrect.
5.2.1 Procedures for Patient Analyzing
1. Press the On/Off button on the left hand side of the analyzer to turn it on.
2. Enter the user name and password manually, and then press .
To enter the user name with the bar code scanner, press first, and then scan the user
name bar code.
Figure 5-1 Enter User Name and Password
3. Press the button for the blood sample type on the Main screen. √ will appear if the button is
selected. The default type is Arterial.
Figure 5-2 the Main Screen