User's Manual
Table Of Contents
- About this Manual
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Table of Contents
- Chapter 1 Safety Guide
- Chapter 2 System Overview
- Chapter 3 Installation Guide
- Chapter 4 Setup
- Chapter 5 Patient Analyzing
- Chapter 6 Quality Control (QC) Tests
- Chapter 7 Data Management
- Chapter 8 Online Update
- Chapter 9 Troubleshooting
- Chapter 10 Cleaning, Care and Maintenance
- Chapter 11 Theory
- Chapter 12 Parameters
- Chapter 13 Warranty and Service
- Appendix 1 Specifications
- Appendix 2 Measurement Ranges
- Appendix 3 Reference Ranges
- Appendix 4 EMC Information
- Appendix 5 FCC Information
- Appendix 6 Order List
i15 Blood Gas and Chemistry Analysis System User Manual Setup
- 52 -
4.3.5 Correlation Factors Setup
Correlation factors allow you to correlate results obtained from the system to results obtained
from another system. The ranges for the slope and offset of each parameter are listed in table 4-3.
Table.4-3 Correlation Factors
Parameter Slope Range Offset Range
pH 1.200 – 0.800 +/- 1.000
pCO
2
1.200 – 0.800
+/- 10.0 mmHg
+/- 1.33 kPa
pO
2
1.200 – 0.800
+/- 40 mmHg
+/- 5.33 kPa
Na
+
1.200 – 0.800 +/-30.0 mmol/L
K
+
1.200 – 0.800 +/- 1.00 mmol/L
Cl
-
1.200 – 0.800 +/- 20.0 mmol/L
Ca
++
1.200 – 0.800 +/- 0.500 mmol/L
Glu 1.200 – 0.800
+/- 50.0 mg/dL
+/- 2.78 mmol/L
Lac 1.200 – 0.800
+/- 4.00 mmol/L
+/- 36.0 mg/dL
Cre 1.200 – 0.800 +/- 30 umol/L
BUN 1.200 – 0.800 +/- 1.8 mmol/L
Hct 1.200 – 0.800 +/- 20 % PCV
1. Press to get access to the Correlation Factors Setup screen.
Figure 4-17 Correlation Factors Setup Screen
2. Edit the values of the slopes and offsets for the parameters whose correlation factors you
want to change.
3. Press OK to accept the changes, and the system will return to the Test Setup screen.