User's Manual
Table Of Contents
- About this Manual
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Table of Contents
- Chapter 1 Safety Guide
- Chapter 2 System Overview
- Chapter 3 Installation Guide
- Chapter 4 Setup
- Chapter 5 Patient Analyzing
- Chapter 6 Quality Control (QC) Tests
- Chapter 7 Data Management
- Chapter 8 Online Update
- Chapter 9 Troubleshooting
- Chapter 10 Cleaning, Care and Maintenance
- Chapter 11 Theory
- Chapter 12 Parameters
- Chapter 13 Warranty and Service
- Appendix 1 Specifications
- Appendix 2 Measurement Ranges
- Appendix 3 Reference Ranges
- Appendix 4 EMC Information
- Appendix 5 FCC Information
- Appendix 6 Order List
i15 Blood Gas and Chemistry Analysis System User Manual Setup
- 46 -
Table 4-1 Patient Information
Parameter Range Resolution Unit
Patient ID 1 - 16 digits N/A N/A
Operator ID 1 - 16 digits N/A N/A
Sample ID 1 - 16 digits N/A N/A
Temperature
14.0 – 44.0 0.1 °C
57.2 – 111.2 0.1 ℉
Sex Male, Female, / N/A N/A
tHb (Total
hemoglobin)
1.0 – 26.0 0.1 g/dL
10 - 260 1 mg/dL
0.6 – 16.1 0.1 mmol/L
FIO
2
0.21 - 1.00 0.01 x.xx
21 - 100 1 %
MCHC (Mean
corpuscular
hemoglobin
concentration)
29.0 - 37.0 0.1 g/dL
290 - 370 1 g/L
RQ (respiratory
quotient)
0.70 - 2.00 0.01 x.xx
Hemoglobin
type
Adult, Children, Newborn N/A N/A
Puncture site
LR/RR/LB/RB/LF/RF/Cord/Scalp/LHF/
RHF/LH/RH/AIC, PA Catheter,
CPB or Other, where:
LR=Left Radial
RR= Right Radial
LB= Left Brachial
RB= Right Brachial
LF= Left Femoral
RF= Right Femoral
LHF= Left Hand Finger
RHF= Right Hand Finger
LH= Left Heel
RH= Right Heel
AIC= Arterial Indwelling Catheter
CPB= Cardiopulmonary Bypass
PA Catheter= PA Catheter
Other= Other
N/A N/A
Bypass Pump Off / Pump On N/A N/A
O
2
mode
Rm air, Mask, T-P, NC, Vent, Bag, Hood or
Other, where:
Rm Air=Room Air
N/A N/A