i15 Blood Gas and Chemistry Analysis System User Manual Setup
- 45 -
4.3.1 QC Lockout Setup
This menu allows you to configure QC lockout. If QC lockout function is enabled, the system
will not report test results for those parameters that have failed quality control tests, and the
results for these parameters will be flagged with xxx. If QC lockout function is disabled, the
system will report test results for those parameters that have failed quality control tests, and the
test results for these parameters will be flagged with ***. The default setting is enabling QC
Lockout function.
Follow the procedures below to set QC Lockout:
1. Press to go to the QC Lockout Setup screen.
Figure 4-11 QC Lockout Setup Screen
2. Select whether to enable QC Lockout function. √ will appear if QC Lockout function is
enabled.
3. Press OK to accept the changes, and the system will return to the Test Setup screen.
4.3.2 Patient Information Setup
This menu allows you to select which parameters will be displayed on the Edit Patient
Information screen during each patient sample test. The default displayed parameters are patient
ID, temperature, FIO
2
and operator ID. The table below lists the parameters for patient
information.