i15 Blood Gas and Chemistry Analysis System User Manual Installation Guide
- 29 -
NOTE:
Avoid tearing the bar code on the foil pouch.
7. Remove the calibrant fluid pack cap, and remove the key by pushing it with your finger in
the direction as shown by the arrow in the picture below.
NOTE:
Avoid pushing the valve when removing the key.
8. Unlock the calibrant fluid pack chamber lock with the key, and then pull the calibrant fluid
pack ejector to open the calibrant fluid pack chamber door.
9. Remove the used calibrant fluid pack from the system. The system will go to the screen for
the next procedure, and the scanner will be turned on automatically.
Figure 3-6 Remove Old Calibrant Fluid Pack
Valve
Key