i15 Blood Gas and Chemistry Analysis System User Manual System Overview
- 14 -
Figure 2-2 Test Cartridge
Table 2-1 Test Cartridge Types
Cartridge
Type
Measured
Parameters
Calculated Parameters
BG8
pH, pCO
2
, pO
2
,
Na
+
, K
+
, Cl
-
, Ca
++
,
Hct
cH
+
, HCO
3
-
act, HCO
3
-
std, BE (ecf), BE (B), BB (B), ctCO
2
,
sO
2
(est), Ca
++
(7.4), AnGap, tHb (est), pO
2
(A-a), pO
2
(a/A),
RI, pO
2
/FIO
2
, cH
+
(T), pH (T), pCO
2
(T), pO
2
(T),
pO
2
(A-a) (T), pO
2
(a/A) (T), RI (T), pO
2
(T)/FIO
2
BG3
pH, pCO
2
, pO
2
cH
+
, HCO
3
-
act, HCO
3
-
std, BE (ecf), BE (B), BB (B), ctCO
2
,
sO
2
(est), pO
2
(A-a), pO
2
(a/A), RI, pO
2
/FIO
2
, cH
+
(T), pH (T),
pCO
2
(T), pO
2
(T), pO
2
(A-a) (T), pO
2
(a/A) (T), RI (T),
pO
2
(T)/FIO
2
Packaging
1. The type of the cartridge is labeled on the test cartridge.
2. Each test cartridge is sealed in a foil pouch containing a strip of desiccant.
3. The bar code on the foil pouch contains information such as the cartridge type, lot number
and expiration date, etc.
4. 25 test cartridges are packaged per box, and 4 boxes are packaged into a shipping carton.
Storage and Usage
Test cartridges should be stored at 4 - 30°C, and should be used at 10 - 31°C with the relative
humidity of 25% - 80% (non-condensing).
Sensor
Arrays
Fluidic
Chamber
Fillport
Electrical
Contacts