User's Manual

Table Of Contents
i15 Blood Gas and Chemistry Analysis System User Manual EMC Information
- 149 -
Appendix 4 EMC Information
Guidance and manufacture’s declaration - electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration electromagnetic emission
The system is intended for use in the electromagnetic environment specified below. The user of
the system should ensure that it is used in such an environment.
Emission test Compliance Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The system
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference to nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The system
is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration - electromagnetic immunity -
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of the system should ensure that it is used in such an environment.
Immunity test
IEC
61326-1&61326-2-6
test level
Compliance level
Electromagnetic
environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±4kV contact
±8kV air
±4kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile.
If floor is covered with
synthetic material, the
relative h
umidity should
be at least 30%.