User's Manual
Table Of Contents
- About this Manual
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Table of Contents
- Chapter 1 Safety Guide
- Chapter 2 System Overview
- Chapter 3 Installation Guide
- Chapter 4 Setup
- Chapter 5 Patient Analyzing
- Chapter 6 Quality Control (QC) Tests
- Chapter 7 Data Management
- Chapter 8 Online Update
- Chapter 9 Troubleshooting
- Chapter 10 Cleaning, Care and Maintenance
- Chapter 11 Theory
- Chapter 12 Parameters
- Chapter 13 Warranty and Service
- Appendix 1 Specifications
- Appendix 2 Measurement Ranges
- Appendix 3 Reference Ranges
- Appendix 4 EMC Information
- Appendix 5 FCC Information
- Appendix 6 Order List
i15 Blood Gas and Chemistry Analysis System User Manual Measurement Ranges
- 147 -
Appendix 2 Measurement Ranges
A2.1 Measurement Ranges for Measured Parameters
Parameter Measurement Range
pH 6.500 - 8.000
pO
2
10 - 700mmHg
pCO
2
10 - 150mmHg
K
+
2.0 - 9.0mmol/L
Na
+
100 - 180mmol/L
Cl
-
65 - 140mmol/L
Ca
++
0.25 - 2.50mmol/L
Hct 10 - 75 %PCV
A2.2 Measurement Ranges for Calculated Parameters
Parameter Measurement Range
cH
+
10.0 - 316.2nmol/L
cH
+
(T) 10.0 - 316.2nmol/L
pH(T) 6.500 - 8.000
pCO
2
(T) 10.0 - 150.0mmHg
pO
2
(T) 10 - 700mmHg
HCO
3
-
act 1.0 - 99.9mmol/L
HCO
3
-
std 1.0 - 99.9mmol/L
BB(B) 13.1 - 82.4mmol/L
BE(B) (-30.0) - (+30.0)mmol/L
BE(ecf) (-30.0) - (+30.0)mmol/L
ctCO
2
1 - 100mmol/L
Ca
++
(7.4) 0.23 - 2.70mmol/L
sO
2
(est) 1% – 100%
AnGap (-10) - (+99)mmol/L
tHb(est) 3.4 - 25.5g/dL
pO
2
(A-a) 0 - 733mmHg
pO
2
(A-a)(T) 0 - 733mmHg
pO
2
(a/A) 0.00 - 1.00
pO
2
(a/A)(T) 0.00 - 1.00
RI 0.00 - 20.00
RI(T) 0.00 - 20.00
pO
2
/FIO
2
250 - 700mmHg
pO
2
(T)/FIO
2
250 - 700mmHg