i15 Blood Gas and Chemistry Analysis System User Manual Specifications
- 146 -
A1.5 Rechargeable Battery
Type Rechargeable lithium battery
Working Time
When the battery is fully charged, the system can analyze and print
50 samples.
Necessary Charge time
No more than 8 hours
Rated Capacity 4200mAh
Rated voltage 14.8V
Charge Mode Constant voltage/current
Charge Current (standard)
0.2C
5
A
Charge Temperature 0°C - +50°C (+32°F - +122°F)
Operating Temperature -20°C - +60°C (-4°F - +140°F)
Storage
Short Term (within 1 month): -20°C - +60°C (-4ºF - +140ºF)
Middle Term (within 3 months): -20°C - +45°C (-4ºF - +113ºF)
Long Term (within 1 year): -20°C - +20°C (-4ºF - +68ºF)
During storage, recharge the battery at least every six months.
Cycle Life
300 times
A1.6 Safety Specifications
Comply with
EN 61010-1: 2001, EN 61010-2-101: 2002,
EN 61326-1: 2006, EN 61326-2-6: 2006
Degree of protection against
harmful ingress of water
Ordinary equipment (Sealed equipment without water proof)
Degree of safety in presence
of flammable gases
Equipment not suitable for use in presence of flammable gases
EMC CISPR 11 Group 1, Class A
Electric shock Class I
Mode of operation Continuous