i15 Blood Gas and Chemistry Analysis System User Manual Specifications
- 145 -
Humidity 25% - 80% (non-condensing)
Ambient
Pressure
70 - 106.6kPa
(525 - 800mmHg)
Transport
Temperature 2 - 30°C
Humidity 25% - 93% (non-condensing)
Ambient
Pressure
40 - 106.6kPa
(300 - 800mmHg)
Storage
Temperature 2 - 8°C (avoid freezing)
Humidity 25% - 93% (non-condensing)
Ambient
Pressure
40 - 106.6kPa
(300 - 800mmHg)
A1.2 Analyzer Specifications
Size 315*238*153mm (length ×width ×height)
Weight 3.8Kg (rechargeable battery included, adaptor not included)
LCD Screen
7 inch, 800*480, Color TFT
Power
Voltage 100V - 240V~
Input Current 1.2A - 0.5A
Frequency 50Hz/60Hz
Rechargeable
Battery
14.8VDC/4200mAh
USB Interface 4 USB Interfaces
Serial Port DB9
Bar Code Scanner embedded
A1.3 Performance Specifications
Sample Volume 110uL
Test Time It takes no more than 70s from sampling to display of test results.
A1.4 Printer
Printer Built-in thermal printer, thermal printer paper
Paper Width 48mm