i15 Blood Gas and Chemistry Analysis System User Manual Specifications
- 144 -
Appendix 1 Specifications
A1.1 Environment Requirements
Analyzer
Usage
Temperature 10 - 31°C
Humidity 25% - 80% (non-condensing)
Ambient
Pressure
70 - 106.6kPa
(525 - 800mmHg)
Transport and
Storage
Temperature -20 - 60°C
Humidity 25% - 93% (non-condensing)
Ambient
Pressure
40 - 106.6kPa
(300 - 800mmHg)
Test
Cartridge
Usage
Temperature 10 - 31°C
Humidity 25% - 80% (non-condensing)
Ambient
Pressure
70 - 106.6kPa
(525 - 800mmHg)
Transport
Temperature 4 - 37°C
Humidity ≤ 93% (non-condensing)
Ambient
Pressure
40 - 106.6kPa
(300 - 800mmHg)
Storage
Temperature 4 - 30°C
Humidity ≤ 93% (non-condensing)
Ambient
Pressure
40 - 106.6kPa
(300 - 800mmHg)
Calibrant
Fluid Pack
Usage
Temperature 10 - 31°C
Humidity 25% - 80% (non-condensing)
Ambient
Pressure
70 - 106.6kPa
(525 - 800mmHg)
Transport and
Storage
Temperature 2 - 8°C (avoid freezing)
Humidity ≤ 93% (non-condensing)
Ambient
Pressure
65 - 106.6kPa
(489 - 800mmHg)
Controls Usage
Temperature 10 - 31°C