User's Manual

Table Of Contents
i15 Blood Gas and Chemistry Analysis System User Manual Cleaning, Care and Maintenance
- 114 -
Figure 10-2 Remaining Time
6. Clean the screen with the lint-free cloth.
7. Repeat steps 4-6 to disinfect the screen with the lint-free cloth after the screen is thoroughly
dry.
NOTE:
Clean the screen prior to disinfecting it.
Make sure that the screen is thoroughly dry before disinfecting it.
8. Press Return twice to return to the Main screen.
10.1.3 Cleaning the Printer Head
Dirty and soiled thermal printer heads will decrease the printing definition, so printer heads
should be cleaned at least once a month.
Open the printer casing and remove the printer paper. Clean the printer head gently with a clean
soft cloth dampened in 75% alcohol. For stubborn stains, soak the stain with a little alcohol first
and wipe it off with a clean soft cloth. After air drying, load the printer paper and shut the printer
casing.
1 Prevent the detergent from seeping into the system while cleaning. Do not immerse
the analyzer in liquid under any circumstances.
CAUTION
2 Do not clean the system with abrasive fabric.