i15 Blood Gas and Chemistry Analysis System User Manual Cleaning, Care and Maintenance
- 113 -
10.1.2 Cleaning and Disinfecting the Screen
Clean and disinfect the screen to remove dust, splatters, blood, etc. Policies regarding the
cleaning and disinfecting intervals are at the discretion of your individual institution. Follow the
steps below to clean and disinfect the screen:
1. Dampen a lint-free cloth with 0.5% sodium hypochlorite solution.
NOTE:
The cloth should be wet but not dripping.
2. On the Main screen, press to go to the Setup screen.
3. Press on the System Setup screen.
4. Select the desired time period during which the system will have no response when you
touch the screen. There are four options: 30 seconds, 1 minute, 2 minutes and 5 minutes. The
default is 30 seconds.
Figure 10-1 Select Clean Time
5. Press Clean Screen. The system will display the screen below: