User's Manual

Table Of Contents
i15 Blood Gas and Chemistry Analysis System User Manual Troubleshooting
- 111 -
28 Fluids have been spilled on the screen.
Clean and disinfect the screen following the
user manual instructions.
29 The USB scanner does not work smoothly.
Check the scanner type. I
f it is the type
recommended by EDAN, remove the
scanner from the system, and insert it again.
Set the scanner, and scan a bar code again.
30 Forget the user name or password.
F
or an operator, contact the administrator
for assistance. For an administrator, contact
EDAN or its authorized distributors for
assistance.
31 The lithium battery can not be charged.
Install the lithium battery following the user
manual instructions. Connect the system to
AC power. T
he system will be charged
automatically. If it still can not be charged,
please contact EDAN or its authorized
distributors for assistance.
32
There are fluids flowing from the bottom of
the analyzer.
M
ove the analyzer horizontally, and clear
the fluids. Remove the calibrant fluid pack
from the analyzer and check whether it is
damaged. I
f it is not damaged, perform a
simulator test and a control test to ensure the
proper performance of the system.
33
The prompt
Ambient Temperature Out of
Rangeappears.
C
heck that the ambient temperature is
within 10 - 31°C, and that the air vents are
not obstructed. Reboot the system. I
f the
problem continues, please contact EDAN or
its authorized distributors for assistance.
34 The prompt “Heating Abnormallyappears.
Please contact EDAN or its authorized
distributors for assistance.
35
The prompt Abnormal Cam Location
appears.
Please contact EDAN or i
ts authorized
distributors for assistance.
36
The date setup can not be saved after you
shut down the system.
P
lease contact EDAN or its authorized
distributors for assistance.
37 The system makes abnormal sounds.
P
lease contact EDAN or its authorized
distributors for assistance.