i15 Blood Gas and Chemistry Analysis System User Manual Troubleshooting
- 109 -
Wellappears.
please contact EDAN or its authorized
distributors for assistance.
8 The prompt “Cartridge Expiredappears.
Verify that the current date is correct. If the
prompt appears again, check that the test
cartridge has not expired. If it has expired,
use a new test cartridge that has not expired.
9
The prompt
Calibrant is Not Aspirated
Properly” appears.
P
erform the test again with a new test
cartridge. I
f the prompt appears again,
please contact EDAN or its authorized
distributors for assistance.
10
The prompt Air is Not Aspirated Properly
appears.
P
erform the test again with a new test
cartridge. I
f the prompt appears again,
please contact EDAN or its authorized
distributors for assistance.
11
The prompt Sample Error. Maybe Clots are
Detected in Sampleappears.
Check that there are no clots in the blood
sample. If there are clots, discard it and
collect samples again.
12 The prompt “Insertion Error” appears.
V
erify that a simulator is inserted during a
simulator test and a test cartridge is inserted
during a sample test.
13
The prompt
Calibrant Pack Expired,
Confirm Date Setup and Calibrant Packs
Expiration Dateappears.
Verify that the current date is correct. If the
prompt appears again, check that the
calibrant fluid pack has not expired. If it has
expired, replace the calibrant fluid pack
following the user manual instructions.
14 The prompt “Remaining Days: 0appears.
T
he calibrant fluid pack has expired.
R
eplace the calibrant fluid pack following
the user manual instructions.
15
The prompt Calibrant Pack Depleted
appears.
The calibrant fluid pack has been depleted.
Replace the calibrant flu
id pack following
the user manual instructions.
16 The prompt “Current Date Errorappears.
The date is earlier than the manufacture date
of the calibrant fluid pack. Ensure that the
current date is correct.
17
The prompt
Calibrant Pack Improper
Removal. Please Replace Calibrant Pack
appears.
The calibrant fluid pack
is removed from
the system improperly. Replace the calibrant
fluid pack following the user manual
instructions.