i15 Blood Gas and Chemistry Analysis System User Manual Data Management
- 103 -
Figure 7-8 Change System Password
4. Press OK to accept the changes, and press OK in the pop up dialog box. The system will
go to the User Login screen.
NOTE:
The default factory system password is 123456.
Only administrators can change the system password (Even engineers from the
manufacturer can not change it, although they can retrieve the password to the
default factory setting).
7.2.2 Diagnosis Database
The diagnosis database displays diagnosis data for the most recent month by default, and it can
store up to 10000 data entries. The following operations can be performed in the diagnosis
database: exporting diagnosis data to a removable disk (such as a USB drive), viewing the details
of diagnosis data, searching for and printing diagnosis data, etc.
NOTE:
Only service engineers and engineers from the manufacturer can get access to the
diagnosis database.
7.2.3 Events Log Database
The events log database records the following information: User Login and Logout, changing
time and date, changing slopes and offsets, changing patient information, deleting data stored in
databases, replacing a calibrant fluid pack. It can store up to 10000 data entries. The system
displays 50 data entries on every page. Press Prev. and Next to page through the screens of the
displayed data entries. You can export, search for and print the events with the events log
database.