i15 Blood Gas and Chemistry Analysis System User Manual Data Management
- 102 -
Figure 7-7 Change Operator Password
4. Press OK to save the changes, and press OK in the pop up dialog box. The system will
go to the Security Database screen.
NOTE:
The password is case sensitive, and should be 4 - 16 characters long (including both
letters and numbers).
7.2.1.4 Deleting Security Data
1. Select the desired security data.
2. Press Delete, and press OK in the pop up message.
NOTE:
If no security data is selected before pressing Delete, all of the data stored in the security
database will be deleted.
7.2.1.5 Changing System Password
1. Press Administrator on the Security Database screen.
2. Press Edit.
3. Enter the current system password, enter the new system password and re-enter the same
password.