i15 Blood Gas and Chemistry Analysis System User Manual Data Management
- 101 -
7.2.1.2 Adding an Operator
1. On the Security Database screen, press Add.
2. Enter the operator ID, password and re-enter the same password.
Figure 7-6 Add Operator
3. Press OK to save the changes, and press OK in the pop up dialog box. The system will
go to the Security Database screen.
NOTE:
The operator ID is case sensitive, and should be 1 - 16 characters long (including
both letters and numbers). "admin/demo/service/edan", or upper and lower case
combinations of these words are not permitted.
The password is case sensitive, and should be 4 - 16 characters long (including both
letters and numbers).
7.2.1.3 Changing Operator Password
1. Select the operator data you want to edit.
2. Press Edit.
3. Enter the password, and re-enter the same password.