i15 Blood Gas and Chemistry Analysis System User Manual Quality Control (QC) Tests
- 94 -
6. The system automatically displays the test result upon the completion of the test.
Figure 6-31 External Simulator Test Result
7. View the result.
NOTE:
Upon the completion of the test, the message Please Remove Cartridgewill be
displayed on the status bar at the bottom of the screen, and the simulator will be
ejected.
If the system passes the external simulator test, it can be used to analyze samples.
If the system fails the external simulator test, perform the test again or try another
simulator. If it passes the test, it can be used to analyze samples. If it fails again,
contact EDAN or its authorized distributors for assistance.
8. Press Home to go to the Main screen.
9. Remove the external simulator from the system, and put it back.
NOTE:
Never remove the external simulator until the test is complete.
6.3.2 Simulator Database
The simulator database displays all of the simulator test data by default, and it can store up to
2000 data entries. The system displays 50 data entries on every page. Press Prev. and Next to
page through the screens of the displayed data entries. When 80% of the space is occupied, the