i15 Blood Gas and Chemistry Analysis System User Manual Quality Control (QC) Tests
- 93 -
3. On the Main screen, press to go to the Quality Control screen, and then press .
4. Gently insert the external electronic simulator into the cartridge port, and press down to
ensure it clicks into place. If the electronic simulator is inserted properly, the system will
automatically perform a simulator test.
Figure 6-29 Insert External Simulator
NOTE:
Avoid touching the contact pads.
Never touch the simulator after inserting it into the system.
5. The system automatically starts to perform an external simulator test.
Figure 6-30 Measurement in Progress