User's Manual

F Series Fetal & Maternal Monitor User Manual Abbreviation
- 168 -
A4.6.2 Test of Wireless Probe (FTS-3)
Acoustic Output Reporting Table
Operating Mode: PW mode
Working Frequency: 1.0MHz
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-scan
Aaprt1 Aaprt>1
Global Maximum Index Value 0.01087 0.002949 0.01939 N/A
Associated
Acoustic
Parameters
P
r.a
(MPa) 0.01087
P (mW) 3.496 N/A
Min of [P
a
(Z
s
),I
ta.a
(Z
s
)] (mW) 0.62
Z
s
(cm) 12.50
Z
bp
(cm) 5.188
Z
b
(cm) 11.95
Z at max I
pi.a
(cm) 12.10 12.10
deq(Z
b
) (cm) 1.77
f
awf
(MHz) 1.00 1.00 1.00 N/A
Dim of Aaprt
X (cm)
3.46 3.46 N/A
Y (cm) 3.46 3.46 N/A
Other
Information
td (usec) 90.65
prr (Hz) 2000
P
r
at max I
pi
(MPa) 0.01743
Deq at max I
pi
(cm) 1.74
I
pi.3
at max MI (W/cm
2
) 0.003412
Focal
Length
Flx (cm)
N/A
Fly (cm) N/A
Operating
Control
Conditions
Focus(mm) Fixed
Depth(mm) Fixed
Freq(MHz) 1.0