User's Manual

PADECG User Manual EMC Information
- 42 -
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1kV for input/output
lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
1 kV line to line
2 kV line to ground
Not applicable
Not applicable
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
30A/m
30A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T;
0,5 cycle
At , 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles )
Single phase: at 0°
0 % U
T
; 250/300
cycle
Not applicable
Not applicable
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
The PADECG is intended for use in the electromagnetic environment specified below. The
customer or the user of the PADECG should assure that it is used in such an environment.
Immunity
test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the