Time Constant:
≥3.2 s
Frequency Response:
0.05Hz ~ 150Hz
Gain:
2.5, 5, 10, 20 (mm/mV) (±5%)
Speed:
5mm/s, 12.5mm/s, 25mm/s or 50mm/s
Input Impedance:
≥20MΩ (10Hz)
Input Circuit Current:
≤0.05μA
Input Voltage Range
<±5mVp-p
Calibration Voltage:
1mV±2%
DC Offset Voltage:
±5 00mV
Minimum Amplitude:
20 μVp-p
Noise:
≤15μVp-p
Recovery Time After Defibrillation
<5 s
CMRR
≥100 dB
Pacemaker Detection (Single Channel Detection)
Amplitude
±2 mV to ± 500 mV
Width
0.1 ms to 2.0 ms
Sampling Frequency
10,000Hz, rhythm lead
DX12(iOS) Transmitter Bluetooth
Transmitting Frequency
2402 Hz ~ 2480Hz
Frequency Band
2402 Hz ~ 2480Hz
Modulation Type
FHSS, GFSK, DPSK, DQPSK
Transmitting Power
≥ 0 dBm
NOTE: Operation of the equipment below the minimum amplitude may cause inaccurate
results.