User's Manual

PADECG User Manual Hint Information
- 30 -
Chapter 5 Hint Information
Hint information and the corresponding causes provided by the system are listed as follows.
Table 5-1 Hint Information and Causes
Hint Information
Causes
Lead off: X
Electrodes fall off the patient or the patient
cable falls off the DX12(iOS) Transmitter.
Lead wire reversal
The limb leads are reversed.
Order server connection fails!
The order server is not enabled or the network
connection is abnormal.
FTP connection fails!
The FTP server is not enabled or the network
connection is abnormal.
Battery of sampling device is weak,
please change the battery after the test!
Battery of DX12(iOS) Transmitter is low.
Battery is weak, and the sampling device
will be powered off!
Battery of DX12(iOS) Transmitter runs out.