User's Manual
SE-1010 PC ECG User Manual Technical Specifications
- 159 -
Filter
Work Frequency
DFT Filter: weak/strong
LOWPASS Filter: 25 Hz / 35 Hz / 45 Hz / 75 Hz / 100 Hz /
150 Hz
CMRR
DP12 ECG sampling box≥110 dB
DX12 transmitter≥100 dB
Pacemaker Detection
Amplitude
DP12 ECG sampling box: ±2 to ±700 mV
DX12 transmitter: ±2 to ±500 mV
Width 0.1 to 2.0 ms
Sampling Frequency 10,000 /sec/channel
NOTE: Test the accuracy of input signal reproduction according to the methods
described in clause 4.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001/(R) 2007, and
the result complies with clause 3.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001/(R)
2007.