SE-1010 PC ECG User Manual Technical Specifications
- 158 -
Paper Speed: 5 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s(±3%)
Record message:
Date, Time, Printing Speed, Filter, Symbol, Heart Rate,
Patient ID, Sex, Age, Lead Mark, Lead Wave, Average
Template Wave or Rhythm Lead Wave, Measurement Result
and Interpretation Information Result (option) etc.
Channel: 3 / 6 / 12 channels, auto baseline adjustment
HR Recognition
Technique: Peak-peak detection
HR Range: 30 BPM ~300 BPM
Accuracy: ±1 BPM
Memory
Memory: Storage amount depends on PC machine
ECG Sampling Box Performance
Leads Mode: 12 standard leads/ Cabrera leads
Acquisition Mode: simultaneously 12 leads
Sample Frequency:
DP12 ECG sampling box: 1,000 /sec/channel
DX12 transmitter: 10,000 /sec/channel (sampling)
500 /sec/channel (analysis)
A/D Resolution:
DP12 ECG sampling box: 24 bits
DX12 transmitter: 18 bits
Time Constant: ≥3.2 s
Frequency Response: 0.05 Hz ~ 150 Hz (-3 dB)
Gain: 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV
Input Impedance:
DP12 ECG sampling box≥50 MΩ (10Hz)
DX12 transmitter≥20 MΩ (10Hz)
Input Circuit Current: ≤0.05 µA
Input Voltage Range <±5 mVpp
Calibration Voltage: 1 mV± 2%
DC Offset Voltage
DP12 ECG sampling box: ±600mV
DX12 transmitter: ±500mV
Noise:
DP12 ECG sampling box≤12.5µVp-p
DX12 transmitter≤15µVp-p