User's Manual

SE-1010 PC ECG User Manual Technical Specifications
- 156 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1,
IEC/EN60601-2-25, ANSI/AAMI EC11, IEC/EN60601-2-51
Anti-electric-shock type: Class ІI
Anti-electric-shock
degree:
Type CF with defibrillation-proof
Degree of protection
against harmful ingress of
water:
Ordinary equipment (Sealed equipment without liquid proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of
application in the presence
of flammable gas:
Equipment not suitable for use in the presence of flammable
gas
Working mode: Continuous operation
EMC: Group 1, Class A
Patient Leakage
Current:
NC <10µA (AC) / <10µA (DC)
SFC <50µA (AC) / <50µA (DC)
Patient Auxiliary
Current:
NC <10µA (AC) / <10µA (DC)
SFC <50µA (AC) / <50µA (DC)
A1.2 Environment Specifications
Transport & Storage Working
Temperature:
DP12 ECG sampling box:
-40ºC
(-40ºF) ~ +55ºC
(+131ºF)
+5ºC (+41ºF) ~ +40ºC (+104ºF)
DX12 ECG sampling box:
-20 (-4ºF)~+55 (+131ºF)
Relative Humidity:
25%~93%
Non-Condensing
25%~80%
Non-Condensing
Atmospheric Pressure: 700hPa ~1060hPa 860hPa ~1060hPa