SE-1010 PC ECG User Manual Configuring the System
- 135 -
9.4.3 Setting Rhythm Lead
The rhythm lead can be one of 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5,
or V6.
When the printing mode is set to 3×4+1 or 6×2+1, the rhythm lead selected in the Rhythm1 list
box will be printed out.
When the printing mode is set to 3×4+3, 3 rhythm leads selected in the Rhythm1, Rhythm2 and
Rhythm3 list boxes will be printed out.
9.4.4 Defining Printing Format
1. The paper form can be set to landscape or portrait.
2. Set Sequence to sequential or synchronous.
When Sequence is set to sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the previous lead group.
When Sequence is set to synchronous, all leads are printed simultaneously. The start time of
each group is the same.
3. Set the paper size to A4 or Letter.
4. Set the Report Hint to Report confirmed by: or Unconfirmed Report.
5. Set the Baseline Adjustment to OFF, Horizontal or Auto.
6. Select Auto gain change, and the gain will be changed automatically and Auto baseline will
be changed to Horizontal automatically.
7. Select Color and the background grid of report will be printed in color.
Deselect Color, and the background grid of report won’t be printed in color.
NOTE: If the printing color is set to color, but a black-and-white printer is used, the
report printed will be illegible.
8. Select Background grid, and the background grid will be printed in the report.
Deselect Background grid, and the background grid won’t be printed in the report.