User's Manual

Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 66 -
7.3.1.2. Strip Doppler Generic Measurements
The Acclarix AX8 supports the following types of generic strip Doppler measurements:
Caliper
Trace
Auto Trace
HR (Heart Rate)
RI (Resistive Index)
Caliper: The Caliper measurement can provide a wide range of results.
To use the Caliper measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If Caliper is not the default measurement, select Caliper from the touch screen.
3. Move the caliper to the desired location. At this point the measurement can display the
following results (the actual results shown will depend on the preset):
Vel
PG
4. If you want a caliper pair, press the Set key to place the caliper and add the second half of
the caliper pair. At this point the measurement may display the following results (the
actual results shown will depend on the preset):
V1
V2
RI
S/D
Time
ΔV
Accel
PG1
PG2
PHT
Further presses of the Set key toggle which side of the caliper can be moved.
Pressing the Enter key completes the caliper and allows you to create another tool.
The S/D, PI, and PHT results assume that the first caliper is placed on the peak
systolic point.
Trace: The Trace measurement can provide a wide range of results.
To use the Trace measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If Trace is not the default measurement, select Trace from the touch screen.
3. Trace can be done with a Draw. Draw will follow your movements
4. Move the caliper to the desired start point.