User's Manual
Acclarix AX8 Diagnostic Ultrasound System User Manual Transducer
- 56 -
There are two methods to disinfect the transducer (use the immersion method for the E8-4Q and
L17-7SQ):
Non-immersion:
1. Disconnect the transducer from the system.
2. Prepare the disinfectant solution (75% ethanol or 70% isopropanol).
3. Spray the solution to the transducer interface or wipe it with a soft cloth dampened with the
disinfectant solution.
4. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry
cloth or leave the transducer to air dry.
5. Inspect the transducer to ensure that there is no damage.
Immersion:
1. Prepare the disinfectant solution (Cidex OPA)
2. Place the cleaned and dried transducer in contact with the disinfectant (refers to figure 6-2 for
the contacting area) for the time specified by the disinfectant manufacturer.
3. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry
cloth or leave the transducer to air dry.
4. Inspect the transducer to ensure that there is no damage.
.
Figure 6-2 Depth of the Transducer Immerged into Disinfectant
WARNING
1. Do not immerse the transducer connector. If the cable connector is immersed, do not
plug the connector into the system. Rinse the connector under running water and dry
it thoroughly. If necessary, contact EDAN for service.
C5-2XQ/
C5-2Q
P5-1XQ
L17-7SQ
L17-7HQ
L10-4Q/
L12-5Q
Disinfectant
E8-4Q
MC8-4Q
C5-2MQ*