5.6.5. 4D Live Volume
4D Live volume continuously sweeps and displays successive 3D volumes. During live 4D
imaging the following controls are available: activate any window of A/B/C/3D,
Single/Dual/Quad,VOI, Threshold, Bright, Contrast, 3D Smooth/ MPR Smooth, 3D Tint/ MPR
Tint, Rotation, Rotation X, Rotation Y, Rotation Z , Zoom and Gain in B mode.
Figure 5-6 4D Volume Sweeping Image
As shown in figure above, image in the left is A slice and image in the right is the live 4D volume
rendering.
5.6.6. 4D Cine
During the process of 4D live volume imaging; press the Freeze button on the control panel to
enter the 4D cine mode. A cine replay scroll bar appears below the images. Use the trackpad to
replay the reset 3D images. For the selected frame, user can execute all operations that are
available 3D review status. Press Freeze again to return to the live image.
Figure 5-7 4D Cine
Touch screen controls are the same as that of static 3D except for Cine button.
Table 5-10 4D Touch Screen
Name
Control
Description
cine