User's Manual
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
- 51 -
Table 5-9 Multi-slice Mode Touch Screen
5.6.4. 4D Volume Acquisition
Press 3D/4D button on the touch screen in B-mode to enter pre-3D, select 4D, select Baby Spine
or Baby Face fist and use the ROI box to set the region. Only ROI will be stored during the
volume sweep. Use the trackpad to set the size and location of the ROI box. Press Set to confirm.
Use the trackpad to set the edge of VOI (Volume of interest) and press Start button on the touch
screen or the Enter button on the trackpad to begin sweep.
Distance
Adjust the distance between slices.
Next
When there are more slices than can be displayed at once this
control cycles through the subset of slices displayed.
Tint
Select multi-slice tint, options of Gold, Sepia, Blue, Ice and
Clear are available.
Smooth
Adjust smoothness of A\B\C slices. Higher filter levels create a
smoother image. Upward presses increase the filter.
Downward presses decrease the filter.
Rotation
X
Rotate the basis slice by the X-axis of the activated window.
Parallel slices change with the rotation operation.
Rotation
Y
Rotate the basis slice by the Y-axis of the activated window.
Parallel slices change with the rotation operation.
Rotation
Z
Rotate the basis slice by the Z-axis of the activated window.
Parallel slices change with the rotation operation.