User's Manual
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 13 -
I/O Ports on the left panel:
Figure 3-2 I/O ports of the system
1. Power supply
5. USB 2.0 port
2. S-Video output port
6. Network port
3. Display port
7. Microphone input port
4. USB 3.0 port
8. Audio output port
Table 3-4 I/O Ports of the System
NOTE
:
If an external monitor is connected to the system via display port, configure the
resolution to 1920x1080, other resolution may cause abnormal ultrasound machine
display.
3.3 Battery Use
The Acclarix AX8 system comes with one lithium-ion battery. When it is fully charged it can run
the system for approximately 60 minutes, depending on use. It is automatically charged when the
system is plugged in.
CAUTION
1. If the system will remain unused for more than one week, charge the battery to at
least 75%capacity, take the battery out and store the system and battery separately.
2. During long term storage the battery should be charged at least once every 3 months
to ensure battery capacity is more than 75%.
1
3
4
5
6
7
2
8