User's Manual
Acclarix AX8 Diagnostic Ultrasound System User Manual EMC Information
- 125 -
Recommended separation distances between portable and mobile
RF communication equipment and the EQUIPMENT or SYSTEM
Recommended separation distances between portable and mobile RF communications
equipment and the Acclarix AX8
The Acclarix AX8 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Acclarix AX8 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Acclarix AX8 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz ~ 80 MHz
Pd 2.1
80 MHz ~ 800 MHz
Pd 2.1
800 MHz ~ 2.5 GHz
Pd 3.2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.