User's Manual

Acclarix AX8 Diagnostic Ultrasound System User Manual Specifications
- 119 -
13 Specifications
13.1 Electrical Safety Classifications
According to the type of
protection against electric
shock
Internally powered equipment,
Class I equipment
According to the degree of
protection against electric
shock
Type BF
According to the degree of
protection against harmful
ingress of liquid
Whole device: IPX0
Console (including touch screen, track pad and hard keys): IPX1
Transducer (not including the transducer connector): IPX7;
Footswitch: IP68
According to the degree of
safety of application in the
presence of a flammable gas
Equipment not suitable for use in the presence of a flammable gas
According to the mode of
operation
Continuous operation
According to the grade of EMC
CISPR 11 Group 1, Class A
Standards Compliance
EN 60601-1:2006
idt IEC 60601-1: 2005 +A1:2012
EN 60601-1-2: 2007
idt IEC 60601-1-2: 2007
IEC/EN 60601-2-37
NEMAUD 2
NEMAUD 3
Imaging Compliance
The AX8 is a portable diagnostic ultrasound system. Compilant
with :
Essential Performance : EN60601-2-37.
2D Imaging: GB10152-2009.
Color Flow Imaging: YY 0767-2009.
Spectral Doppler Imaging: YY 0767-2009.
13.2 Power Supply
Operating Voltage
100 -240 V~
Operating Frequency
50 Hz/60 Hz
AC Input Current
2.5A-1.5A
DC Input Current
19V 7.8A MAX
Lithium battery
Capacity
6150 mAh
Voltage
14.4V
Average working time
1 Hours
Charging time
3 hours with system power-on; 2.5 hours with system power-off
Cycle life
300