User's Manual

Acclarix AX8 Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
- 118 -
WARNING
The console (including touch screen, track pad and hard keys) of system is splash resistant (IPX 1) but
not waterproof. Do not immerse or expose to extended moisture. Splash resistance does not
extend to transducer connectors. Keep connectors dry.
For cleaning agents and disinfectants, refer to section 6.3 Transducer Cleaning and Disinfection
12.4.1.Cleaning the System Surface
To clean the system surface:
1. Switch off the system and unplug it.
2. Wipe the entire exterior surface, including the screen, of the equipment using a soft cloth
dampened with the cleaning solution thoroughly until no visible contaminants remain.
3. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after
cleaning until no visible cleaning agent remains.
4. Dry the system in a ventilated and cool place.
NOTE:
1. Make sure the system is free of gel and any other visible residue.
2. Use a soft dry cloth without chemicals for cleaning. The monitor surface is easily
scratched.
12.4.2.Needle Guide Bracket Cleaning and Sterilization
NOTE:
1. Use proper sterilization technique at all times when performing a biopsy.
2. Ensure that protective gloves are worn.
WARNING
The needle guide bracket kits are not disinfected or sterilized before delivery. The
operators must clean and sterilize the needle guide bracket kits before and after each
use.
Cleaning
Clean the needle guide assembly using a mild detergent prior to sterilizing.
Sterilization
Sterilize the Needle Guide Assembly by autoclaving for at least four minutes at 132
o
C