5.4 Supply voltage connection
Operation of the econ sens3 requires a supply voltage of 100 to 240 VAC. Fig. 6 shows the
connection of the supply input to the network. Connection examples for 1 and 3-phase networks are
found in Chapter 5.10.
When connecting the supply voltage, the following must be noted:
1.) Disconnect device:
An appropriate disconnect device must be installed upstream of the econ sens3 “supply”
connection in order to disconnect the instrument from the current and voltage supply.
This disconnect device must be
Installed close to the instrument and easy for the user to access
Marked as disconnect device for the instrument.
2 Supply / auxiliary connection
Pin No.
Name
Function
1
Connection for functional earth (PE)
2
N
Connection for neutral conductor
3
PRN
Connection for neutral conductor to current
supply
4
PRL
Connection of a phase to current supply
Figure 6: Connection of supply voltage