User manual

5. Managing DE Devices using MC Toolkit Software
5.5. Online configuration
Revision 12 MC Toolkit User Manual 81
May 2013
Table 13 DE Main Menu Procedures
DE Main Menu Procedures
DE Main
Menu
(ST 3000)
ST 800
(STT 3000)
Note:
This message appears if the
<Back button is selected if the
transmitter was set to Output
Mode (in Calibration
procedures), and the Output was
not later cleared.
Various buttons are provided
on the Main Menu to navigate
to different screens.
Save to File allows saving the
current device configuration to
a file. More detail on this under
“Save To File in Online Mode”
Procedure
Device Info
Type field will show
“ST 800” when ST 800
device is connected
Tag ID (r/w)
User ID up to 8 alphanumeric
characters (suggestion: relate
to functional process entities
and/or plant areas).
Type (r)
Manufacturer's device type
identifier (typically, a model
number)
Firmware Version
(r)Manufacturer's Firmware
version identifier
Serial Number (r)
Manufacturer
Scratch Pad (r/w)
Up to 32 alphanumeric
characters (suggestion:
messages to control room
regarding observed/assigned
operational status)